Catastrophe! While slightly moving the NPL tank, a short-circuit happened. I saw sparks and the GC/MS stopped. Everything that was connected to CKT 10 stopped. Strange thing, the fast NOx instrument stopped as well (at least the UPS + fan of H20 pump + laptop). NO flow on fast O3 instrument dropped to 0. I think the line between the NPL tank and the GC/MS touched CKT 10 and that’s how everything happened (see pics below). Jacques and I disconnected everything (computer, GC/MS, UPS) from CKT 10 and connected to CKT 17 instead. GC/MS restarted, fast NOx as well. We will check tomorrow that everything is indeed okay.
The NO flow dropping to 0 on the fast O3 has actually nothing to do with the short-circuit. The needle valve was not sufficiently open. Problem fixed. Keep an eye on this NO flow to make sure it doesn’t drop to 0 again.
Pics of the NPL tank line, CKT 10 and floor below CKT 10.